Overslaan en naar de inhoud gaan
Publicaties | Personen | Instituten | Projecten
[ meld een fout in dit record ]mandje (0): toevoegen | toon Print deze pagina

Solid microelectrodes for in situ voltammetric measurements
Pizeta, I.; Billon, G.; Fischer, J.-C.; Wartel, M. (2003). Solid microelectrodes for in situ voltammetric measurements. Electroanalysis 15(17): 1389-1396. https://dx.doi.org/10.1002/elan.200302669
In: Electroanalysis. Wiley-VCH: Weinheim. ISSN 1040-0397; e-ISSN 1521-4109, meer
Peer reviewed article  

Beschikbaar in  Auteurs 

Trefwoord
    Marien/Kust

Auteurs  Top 
  • Pizeta, I., meer
  • Billon, G.
  • Fischer, J.-C., meer
  • Wartel, M., meer

Abstract
    Solid Au and Ag microelectrodes for both,model,and in situ voltammetric measurements, have been designed andtested. Amalgamation procedure and the parameters of measurement procedure are discussed. Differences betweengold and silver electrodes were studied. Silver electrodes showed more stable behavior and were used prevalently for calibration and in situ measurements.Calibration of a 30-µm diameter Ag electrode for Mn(II), Fe(II), I(-I) andS(-II) showed good linearity with reasonable confidence intervals, necessary for quantification of the signalsobtained in situ, with average sensitivity of 0.05, 0.01, 0.16 and 1.10 nA/(µmol L-1) and minimal detection limit of 3.5, 28, 0.3 and 0.05 µmol L-1, respectively. Additional checking of the electrode sensitivity with the reduction peaks of oxygen is proposed.

Alle informatie in het Integrated Marine Information System (IMIS) valt onder het VLIZ Privacy beleid Top | Auteurs